A source of experimental underestimation of aerosol bolus deposition

Sylvia Verbanck, Carl Darquennes, G. Kim Prisk, Walter Vincken, M. Paiva

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)

Abstract

Abstract

We examined the measurement error in inhaled and exhaled aerosol concentration resulting from the bolus delivery system when small volumes of monodisperse aerosols are inspired to different lung depths. A laser photometer that illuminated similar to 75% of the breathing path cross section recorded low inhaled bolus half-widths (42 ml) and negative deposition values for shallow bolus inhalation when the inhalation path of a 60-ml aerosol was straight and unobstructed. We attributed these results to incomplete mixing of the inhaled aerosol bolus over the breathing path cross section, on the basis of simultaneous recordings of the photometer with a particle-counter sampling from either the center or the edge of the breathing path. Inserting a 90 degrees bend into the inhaled bolus path increased the photometer measurement of inhaled bolus half-width to 57 mi and yielded positive deposition values. Dispersion, which is predominantly affected by exhaled bolus half-width, was not significantly altered by the 90 degrees bend. We conclude that aerosol bolus-delivery systems should ensure adequate mixing of the inhaled bolus to avoid error in measurement of bolus deposition.
Original languageEnglish
Pages (from-to)1067-1074
Number of pages8
JournalJournal of Applied Physiology
Volume86
Issue number3
Publication statusPublished - Mar 1999

Keywords

  • inspired bolus characteristics
  • aerosol deposition
  • photometer
  • particle counter

Fingerprint

Dive into the research topics of 'A source of experimental underestimation of aerosol bolus deposition'. Together they form a unique fingerprint.

Cite this