Condensation of humidified air in the inflation line of a polyurethane cuff precludes correct continuous pressure monitoring during mechanical ventilation

Herbert Spapen, Walter Moeyersons, Stiers Wim, Geert Desmet, E. Suys

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)

Abstract

Within continuously controlled limits of cuff pressure, an endotracheal tube cuff made of polyurethane (PU) prevents secretion inflow better than a conventional polyvinylchloride cuff. However, the temperature difference between ventilator gas and the air inside a PU cuff produces condensation droplets that accumulate in the cuff inflation canal. We investigated whether condensation influenced continuous cuff pressure monitoring. A PU-cuffed tube was placed into an artificial trachea and connected to a ventilator and test lung. An additional line was inserted at the distal cuff end to directly measure pressure inside the cuff. Methylene blue instillation via the inflation line mimicked condensation. Inspiratory (Pinsp) and expiratory (Pexp) pressures were continuously recorded. Six consecutive experiments were performed comparing pressures at baseline (T0) and at 24 h (T24). Shortly after dye instillation, pressures recorded at the inflation canal became permanently fixed at 25 cmH2O. In contrast, pressures measured directly in the cuff progressively decreased (mean Pinsp 30 ± 3 vs. 20 ± 2 cmH2O and mean Pexp 25 ± 0 vs. 12 ± 2 cmH2O, T0 vs. T24, both P <0.05). Thus, condensation in the inflation line of a PU-cuffed tube renders continuous cuff pressure monitoring unreliable, thereby increasing the risk for microaspiration.
Original languageEnglish
Pages (from-to)949-951
JournalJournal of Anesthesia
Volume28
Issue number6
DOIs
Publication statusPublished - 2014

Keywords

  • polyurethane cuff
  • mechanical ventilation

Fingerprint

Dive into the research topics of 'Condensation of humidified air in the inflation line of a polyurethane cuff precludes correct continuous pressure monitoring during mechanical ventilation'. Together they form a unique fingerprint.

Cite this