TY - JOUR
T1 - Evaluation of the anti-factor Xa chromogenic assay for the measurement of rivaroxaban plasma concentrations using calibrators and controls
AU - Samama, Meyer Michel
AU - Contant, Genevieve
AU - Spiro, Theodore E
AU - Perzborn, Elisabeth
AU - Guinet, Céline
AU - Gourmelin, Yves
AU - Le Flem, Léna
AU - Rohde, Gabriele
AU - Martinoli, Jean Luc
AU - Rivaroxaban Anti-Factor Xa Chromogenic Assay Field Trial Laboratories
AU - Jochmans, Kristin
PY - 2012/2
Y1 - 2012/2
N2 - Rivaroxaban is an oral, direct factor Xa inhibitor. Routine coagulation monitoring is not required, but a quantitative determination of rivaroxaban concentrations might be useful in some clinical circumstances. This multicentre study assessed the suitability of the anti-factor Xa chromogenic assay for the measurement of rivaroxaban plasma concentrations (ng/ml) using rivaroxaban calibrators and controls, and the inter-laboratory precision of the measurement. Twenty-four centres in Europe and North America were provided with sets of rivaroxaban calibrators (0, 41, 209 and 422 ng/ml) and a set of rivaroxaban pooled human plasma controls (20, 199 and 662 ng/ml; the concentrations were unknown to the participating laboratories). The evaluation was carried out over 10 days by each laboratory using local anti-factor Xa reagents as well as the centrally provided reagent, a modified STA® Rotachrom® assay. A calibration curve was produced each day, and the day-to-day precision was evaluated by testing three human plasma controls. When using the local anti-factor Xa reagents, the mean rivaroxaban concentrations (measured/actual values) were: 17/20, 205/199 and 668/662 ng/ml, and the coefficient of variance (CV) was 37.0%, 13.7% and 14.1%, respectively. When the modified STA Rotachrom method was used, the measured/actual values were: 18/20, 199/199 and 656/662 ng/ml, and the CV was 19.1%, 10.9% and 10.0%, respectively. The results suggest that, by using rivaroxaban calibrators and controls, the anti-factor Xa chromogenic method is suitable for measuring a wide range of rivaroxaban plasma concentrations (20-660 ng/ml), which covers the expected rivaroxaban plasma levels after therapeutic doses.
AB - Rivaroxaban is an oral, direct factor Xa inhibitor. Routine coagulation monitoring is not required, but a quantitative determination of rivaroxaban concentrations might be useful in some clinical circumstances. This multicentre study assessed the suitability of the anti-factor Xa chromogenic assay for the measurement of rivaroxaban plasma concentrations (ng/ml) using rivaroxaban calibrators and controls, and the inter-laboratory precision of the measurement. Twenty-four centres in Europe and North America were provided with sets of rivaroxaban calibrators (0, 41, 209 and 422 ng/ml) and a set of rivaroxaban pooled human plasma controls (20, 199 and 662 ng/ml; the concentrations were unknown to the participating laboratories). The evaluation was carried out over 10 days by each laboratory using local anti-factor Xa reagents as well as the centrally provided reagent, a modified STA® Rotachrom® assay. A calibration curve was produced each day, and the day-to-day precision was evaluated by testing three human plasma controls. When using the local anti-factor Xa reagents, the mean rivaroxaban concentrations (measured/actual values) were: 17/20, 205/199 and 668/662 ng/ml, and the coefficient of variance (CV) was 37.0%, 13.7% and 14.1%, respectively. When the modified STA Rotachrom method was used, the measured/actual values were: 18/20, 199/199 and 656/662 ng/ml, and the CV was 19.1%, 10.9% and 10.0%, respectively. The results suggest that, by using rivaroxaban calibrators and controls, the anti-factor Xa chromogenic method is suitable for measuring a wide range of rivaroxaban plasma concentrations (20-660 ng/ml), which covers the expected rivaroxaban plasma levels after therapeutic doses.
KW - Anticoagulants
KW - Blood Chemical Analysis
KW - Calibration
KW - Chromogenic Compounds
KW - Europe
KW - Factor Xa Inhibitors
KW - Feasibility Studies
KW - Humans
KW - Morpholines
KW - North America
KW - Observer Variation
KW - Reproducibility of Results
KW - Sensitivity and Specificity
KW - Thiophenes
U2 - 10.1160/TH11-06-0391
DO - 10.1160/TH11-06-0391
M3 - Article
C2 - 22187012
VL - 107
SP - 379
EP - 387
JO - Thrombosis and Haemostasis
JF - Thrombosis and Haemostasis
SN - 0340-6245
IS - 2
ER -