TY - JOUR
T1 - Influence of dabigatran and rivaroxaban on routine coagulation assays. A nationwide Belgian survey
AU - Van Blerk, Marjan
AU - Bailleul, Els
AU - Chatelain, Bernard
AU - Demulder, Anne
AU - Devreese, Katrien
AU - Douxfils, Jonathan
AU - Jochmans, Kristin
AU - Mullier, François
AU - Wijns, Walter
AU - Soumali, Mohamed Rida
AU - Coucke, Wim
AU - Vernelen, Kris
AU - Van de Walle, Philippe
PY - 2015/1
Y1 - 2015/1
N2 - The Belgian national External Quality Assessment Scheme performed a nationwide survey using lyophilised plasma samples spiked with dabigatran or rivaroxaban to demonstrate to the Belgian clinical laboratories how these drugs affect their routine coagulation assays prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen and antithrombin. Virtually all Belgian laboratories performing routine coagulation testing (189/192) participated in the survey. Both, dabigatran and rivaroxaban significantly prolonged the PT and aPTT in a concentration- and reagent-dependent manner. PT reagents were more influenced by rivaroxaban than by dabigatran and aPTT reagents more influenced by dabigatran than by rivaroxaban. Among PT reagents, Neoplastin R® was the most sensitive to rivaroxaban and Innovin® and Thromborel S® the least sensitive. Converting PT results to INR only increased the variability between reagents. Among aPTT reagents, Actin FSL® was the least sensitive to dabigatran while the other aPTT reagents showed slightly higher sensitivities. The presence of dabigatran led to falsely reduced fibrinogen concentrations when measured with a low thrombin concentration reagent. The presence of dabigatran caused an overestimation of the antithrombin level when measured with a thrombin-based activity assay and the presence of rivaroxaban an overestimation of the antithrombin level when measured with a FXa-based assay. Instrument-related differences were found for all tested parameters. In conclusion, this paper provides detailed information on the effect of dabigatran and rivaroxaban on routine coagulation assays as performed with a large number of reagent/instrument combinations.
AB - The Belgian national External Quality Assessment Scheme performed a nationwide survey using lyophilised plasma samples spiked with dabigatran or rivaroxaban to demonstrate to the Belgian clinical laboratories how these drugs affect their routine coagulation assays prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen and antithrombin. Virtually all Belgian laboratories performing routine coagulation testing (189/192) participated in the survey. Both, dabigatran and rivaroxaban significantly prolonged the PT and aPTT in a concentration- and reagent-dependent manner. PT reagents were more influenced by rivaroxaban than by dabigatran and aPTT reagents more influenced by dabigatran than by rivaroxaban. Among PT reagents, Neoplastin R® was the most sensitive to rivaroxaban and Innovin® and Thromborel S® the least sensitive. Converting PT results to INR only increased the variability between reagents. Among aPTT reagents, Actin FSL® was the least sensitive to dabigatran while the other aPTT reagents showed slightly higher sensitivities. The presence of dabigatran led to falsely reduced fibrinogen concentrations when measured with a low thrombin concentration reagent. The presence of dabigatran caused an overestimation of the antithrombin level when measured with a thrombin-based activity assay and the presence of rivaroxaban an overestimation of the antithrombin level when measured with a FXa-based assay. Instrument-related differences were found for all tested parameters. In conclusion, this paper provides detailed information on the effect of dabigatran and rivaroxaban on routine coagulation assays as performed with a large number of reagent/instrument combinations.
KW - rivaroxaban
KW - Dabigatran
KW - external qualitiy assessment
KW - routine coagulation assays
U2 - 10.1160/TH14-02-0161
DO - 10.1160/TH14-02-0161
M3 - Article
C2 - 25231101
VL - 113
SP - 154
EP - 164
JO - Thrombosis and Haemostasis
JF - Thrombosis and Haemostasis
SN - 0340-6245
IS - 1
ER -