Influence of probe application pressure on in vitro and in vivo capacitance (Corneometer CM825) and Conductance (Skicon 22ZX) measurements

Research output: Contribution to journalArticlepeer-review

Abstract

Background/purpose: The measuring probe of the electrical hydration instruments is equipped with a spring system to ensure a constant pressure of probe application on the skin surface. However, it is possible to trigger the measurement at, respectively, a low and a high force of probe application. It is the purpose of this paper to compare the in vitro and in vivo hydration values of the capacitance and conductance method at, respectively, low and high values of application pressure on the considered surface.

Methods: In vitro capacitance and conductance hydration measurements were carried out at a low and a high pressure on a calibration filter pad impregnated with different alcohols and in vivo on different skin sites covering a wide range of hydration values.

Results: The in vitro results were pressure dependent and higher hydration values were noticed when the probes were applied with a high pressure compared with a low pressure.

In vivo, the pressure effect was a function of the skin type and it was more pronounced for dry skin (40%), lower for normal skin (20%) and the lowest for hydrated skin (5–8%).

Conclusion: This study confirms the influence of probe pressure application on the skin surface when carrying out capacitance and conductance hydration measurements.
Original languageEnglish
Pages (from-to)445-450
JournalSkin Research and Technology
Volume17
Issue number4
Publication statusPublished - 2010

Keywords

  • Influence of probe application pressure on in vitr

Fingerprint

Dive into the research topics of 'Influence of probe application pressure on in vitro and in vivo capacitance (Corneometer CM825) and Conductance (Skicon 22ZX) measurements'. Together they form a unique fingerprint.

Cite this