Abstract
BACKGROUND: When the concentrations of 2 or more substances are measured separately, their molar ratios are subject to the additive imprecisions of the different assays. We hypothesized that the cumulative error for concentration ratios of peptides containing a common sequence might be minimized by measuring the peptides simultaneously with a "trefoil-type" immunoassay.
METHODS: As a model of this approach, we developed a dual-label time-resolved fluorescence immunoassay (TRFIA) to simultaneously measure proinsulin, C-peptide, and the proinsulin-C-peptide ratio (PI/C). A monoclonal antibody captures all C-peptide-containing molecules, and 2 differently labeled antibodies distinguish between proinsulin-like molecules and true C-peptide.
RESULTS: The trefoil-type TRFIA was capable of measuring plasma C-peptide and proinsulin simultaneously without mutual interference at limits of quantification of 48 and 8125 pmol/L, and 2.1 and 197 pmol/L, respectively. Within-laboratory imprecision values for the trefoil-type TRFIA ranged between 8.4% and 12% for the hormone concentrations. Unlike the hormone results obtained with separate assays, imprecision did not increase when PI/C was calculated from trefoil assay results (P <0.05). Peptide concentrations were highly correlated with results obtained in individual comparison assays (r2 0.965; P <0.0001). The total error for PI/C obtained with the trefoil-type TRFIA remained 25% over a broader C-peptide range than with separate hormone assays (79-7200 pmol/L vs 590-4300 pmol/L C-peptide). Preliminary data indicate little or no interference by heterophile antibodies.
CONCLUSIONS: The developed trefoil-type TRFIA is a reliable method for simultaneous measurement of proinsulin, C-peptide, and PI/C and provides proof of principle for the development of other trefoil-type multiple-label immunoassays.
METHODS: As a model of this approach, we developed a dual-label time-resolved fluorescence immunoassay (TRFIA) to simultaneously measure proinsulin, C-peptide, and the proinsulin-C-peptide ratio (PI/C). A monoclonal antibody captures all C-peptide-containing molecules, and 2 differently labeled antibodies distinguish between proinsulin-like molecules and true C-peptide.
RESULTS: The trefoil-type TRFIA was capable of measuring plasma C-peptide and proinsulin simultaneously without mutual interference at limits of quantification of 48 and 8125 pmol/L, and 2.1 and 197 pmol/L, respectively. Within-laboratory imprecision values for the trefoil-type TRFIA ranged between 8.4% and 12% for the hormone concentrations. Unlike the hormone results obtained with separate assays, imprecision did not increase when PI/C was calculated from trefoil assay results (P <0.05). Peptide concentrations were highly correlated with results obtained in individual comparison assays (r2 0.965; P <0.0001). The total error for PI/C obtained with the trefoil-type TRFIA remained 25% over a broader C-peptide range than with separate hormone assays (79-7200 pmol/L vs 590-4300 pmol/L C-peptide). Preliminary data indicate little or no interference by heterophile antibodies.
CONCLUSIONS: The developed trefoil-type TRFIA is a reliable method for simultaneous measurement of proinsulin, C-peptide, and PI/C and provides proof of principle for the development of other trefoil-type multiple-label immunoassays.
Original language | English |
---|---|
Pages (from-to) | 1990-1998 |
Number of pages | 8 |
Journal | Clinical Chemistry |
Volume | 54 |
Issue number | 12 |
Publication status | Published - 4 Dec 2008 |
Bibliographical note
Nader RifaiKeywords
- Diabetes
- immunoassays
- proinsulin
- C-peptide
- insulin
- simultaneous
- multiplex
- time resolved fluorescecnce
- trefoil type