Samenvatting
AIM: An ultrasensitive nano UHPLC-ESI-MS/MS method is developed to simultaneously monitor three low-concentration neuromedin-like peptides in microdialysates.
RESULTS: Peptide preconcentration and sample desalting is performed online on a trap column. A shallow gradient slope at 300 nl/min on the analytical column maintained at 35°C, followed by two saw-tooth column wash cycles, results in the highest sensitivity and the lowest carryover. The validated method allows the accurate and precise quantification of 0.5 pM neurotensin and neuromedin N (2.5 amol on column), and of 3.0 pM neuromedin B (15.0 amol on column) in in vivo microdialysates without the use of internal standards.
CONCLUSION: The assay is an important tool for elucidating the role of these neuromedin-like peptides in the pathophysiology of neurological disorders.
Originele taal-2 | English |
---|---|
Pagina's (van-tot) | 605-619 |
Aantal pagina's | 15 |
Tijdschrift | Bioanalysis |
Volume | 7 |
Nummer van het tijdschrift | 5 |
DOI's | |
Status | Published - 2015 |