Applicability of linear and nonlinear retention-time models for reversed-phase liquid chromatography separations of small molecules, peptides, and intact proteins

Eva Tyteca, Jelle De Vos, Nikola Vankova, Petr Cesla, Gert Desmet, Sebastiaan Eeltink

Onderzoeksoutput: Articlepeer review

23 Citaten (Scopus)

Samenvatting

The applicability and predictive properties of the linear solvent strength model and two non-linear retention-time models, i.e., the quadratic model and the Neue model, were assessed for the separation of small molecules (phenol derivatives), peptides, and intact proteins. Retention-time measurements were conducted in isocratic mode and gradient mode applying different gradient times and elution-strength combinations. The quadratic model provided the most accurate retention-factor predictions for small molecules (average absolute prediction error of 1.5%) and peptides separations (with a prediction error of 2.3%). An advantage of the Neue model is that it can provide accurate predictions based on only three gradient scouting runs, making tedious isocratic retention-time measurements obsolete. For peptides, the use of gradient scouting runs in combination with the Neue model resulted in better prediction errors (<2.2%) compared to the use of isocratic runs. The applicability of the quadratic model is limited due to a complex combination of error and exponential functions. For protein separations, only a small elution window could be applied, which is due to the strong effect of the content of organic modifier on retention. Hence, the linear retention-time behavior of intact proteins is well described by the linear solvent strength model. Prediction errors using gradient scouting runs were significantly lower (2.2%) than when using isocratic scouting runs (3.2%).
Originele taal-2English
Pagina's (van-tot)1249-1257
Aantal pagina's9
TijdschriftJournal of Separation Science
Volume39
Nummer van het tijdschrift7
DOI's
StatusPublished - apr 2016

Vingerafdruk

Duik in de onderzoeksthema's van 'Applicability of linear and nonlinear retention-time models for reversed-phase liquid chromatography separations of small molecules, peptides, and intact proteins'. Samen vormen ze een unieke vingerafdruk.

Citeer dit