Samenvatting
Photoplethysmography (PPG) is a widely used technique to extract physiological information non-invasively. Besides other factors, the transimpedance amplifier (TIA) topology choice has an impact on the PPG signal. This study presents a novel quantitative evaluation of seven different resistive TIA topologies in terms of four different Signal Quality Indexes (SQIs): AC amplitude, perfusion index, signal shape quality and signal-to-noise-ratio. Results indicate that a differential TIA configuration is preferred over the single-stage and that while the PD polarity has no impact on the PPG signal quality, the bias choice is important, being the zero bias configuration more beneficial than the reverse bias.
Originele taal-2 | English |
---|---|
Titel | Effect of the Transimpedance Amplifier Topology on the Photoplethysmography Signal |
Uitgeverij | IEEE Xplore |
Pagina's | 1-7 |
Aantal pagina's | 7 |
ISBN van elektronische versie | 978-1-6654-8025-3 |
ISBN van geprinte versie | 978-1-6654-8026-0 |
DOI's | |
Status | Published - 18 okt 2022 |
Evenement | 48th Annual Conference of the IEEE Industrial Electronics Society IECON 2022 : Conference session- Power Electronics & Energy Conversion - Brussels, Brussels, Belgium Duur: 17 okt 2022 → 20 okt 2022 https://iecon2022.org/conference-program/ |
Publicatie series
Naam | IECON Proceedings (Industrial Electronics Conference) |
---|---|
Volume | 2022-October |
Conference
Conference | 48th Annual Conference of the IEEE Industrial Electronics Society IECON 2022 |
---|---|
Verkorte titel | IECON2022 |
Land/Regio | Belgium |
Stad | Brussels |
Periode | 17/10/22 → 20/10/22 |
Internet adres |
Bibliografische nota
Funding Information:This work has been possible thanks to the FWO-Flanders FWOSB106 PhD grant. This work has also been partially supported by the Belgian Development Cooperation through the VLIR-UOS project IUC 2019 Phase 2 with UO.
Publisher Copyright:
© 2022 IEEE.
Copyright:
Copyright 2022 Elsevier B.V., All rights reserved.