8 Citaten (Scopus)

Samenvatting

The applicability of the kinetic plot theory to temperature-programmed gas chromatography (GC) has been confirmed experimentally by measuring the efficiency of a temperature gradient separation of a simple test mixture on 15, 30, 60 and 120m long (coupled) columns. It has been shown that the temperature-dependent data needed for the kinetic plot calculation can be obtained from isothermal experiments at the significant temperature, a temperature that characterizes the entire gradient run. Furthermore, optimal flow rates have been calculated for various combinations of column length, diameter, and operating temperature (or significant temperature). The tabulated outcome of these calculations provide good starting points for the optimization of any GC separation. (C) 2016 Elsevier B.V. All rights reserved.
Originele taal-2English
Pagina's (van-tot)94-100
Aantal pagina's6
TijdschriftJournal of Chromatography A
Volume1450
DOI's
StatusPublished - 10 jun 2016

Vingerafdruk

Duik in de onderzoeksthema's van 'Kinetic plots for programmed temperature gas chromatography'. Samen vormen ze een unieke vingerafdruk.

Citeer dit