Mitral Valve Morphology Assessment: Three-Dimensional Transesophageal Echocardiography Versus Computed Tomography

Miriam Shanks, Victoria Delgado, Arnold Ng, Frank Van Der Kley, Joanne Schuijf, C. Boersma, Nico Van De Veire, Gaetano Nucifora, Matteo Bertini, Albert De Roos, Lucia Kroft, Martin Schalij, Jeroen Bax

Onderzoeksoutput: Articlepeer review

48 Citaten (Scopus)

Samenvatting

Background. Advances in the minimally invasive mitral valve repair techniques increase the demands on accurate and reliable morphologic assessment of the mitral valve using three-dimensional imaging modalities. The present study compared mitral valve geometry measurements obtained by three-dimensional transesophageal echocardiography (TEE) to those obtained with multidetector row computed tomography (MDCT) used as a standard reference.
Methods. Clinical preoperative MDCT and intraoperative three-dimensional TEE were performed in 43 patients (mean age 81.0 +/- 7.7 years) considered for transcatheter valve implantation procedure. Various measurements of mitral valve geometry were obtained from three-dimensional TEE datasets using mitral valve quantification software, and compared with those obtained from MDCT images using multiplanar reformation planes.

Results. Moderate and severe mitral regurgitation was present in 48.9% of patients. There was good agreement in mitral valve geometry measurements between three-dimensional TEE and MDCT without significant overestimation or underestimation and tight 95% limits of agreement. For linear dimensions, angles and areas, the 95% limits of agreement were less than 1 cm, less than 15 degrees, and less than 2 cm(2), respectively. In addition, the intraclass correlation coefficients were more than 0.8 for all parameters. Finally, the measurements were highly reproducible, with low intraobserver and interobserver variability (nonsignificant overestimation or underestimation and narrow 95% limits of agreement).

Conclusions. The present study demonstrates the accuracy and clinical feasibility of the assessment of the mitral valve geometry with three-dimensional TEE that is comparable to the MDCT measurements. Three-dimensional TEE and MDCT provide accurate and complementary information in the evaluation of patients with mitral valve disease. Its potential incremental clinical value in the field of transcatheter mitral repair procedures needs further assessment in the future studies. (Ann Thorac Surg 2010;90:1922-9) (C) 2010 by The Society of Thoracic Surgeons
Originele taal-2English
Pagina's (van-tot)1922-1929
Aantal pagina's8
TijdschriftAnnals of Thoracic Surgery
Volume90
Nummer van het tijdschrift6
StatusPublished - nov. 2010

Vingerafdruk

Duik in de onderzoeksthema's van 'Mitral Valve Morphology Assessment: Three-Dimensional Transesophageal Echocardiography Versus Computed Tomography'. Samen vormen ze een unieke vingerafdruk.

Citeer dit