Doorgaan naar hoofdnavigatie Doorgaan naar zoeken Ga verder naar hoofdinhoud

Optimization and validation of a LaBr3 (Ce) detector model for use in Monte Carlo simulations

M.M. Hasan, T. Vidmar, J. Rutten, L. Verheyen, J. Camps, M. Huysmans

Onderzoeksoutput: Articlepeer review

13 Citaten (Scopus)
177 Downloads (Pure)

Samenvatting

A reliable detector model is needed for Monte Carlo efficiency calibration. A LaBr3(Ce) detector model was optimized and verified using different radioactive sources (241Am, 133Ba, 137Cs, 60Co and 152Eu) and geometries (point, extended and surface). PENELOPE and MCNP were used for Monte Carlo simulations. A good agreement was observed between simulated and experimental full energy peak efficiencies (FEPE) as their mean relative difference was 2.84% ± 1.93% and 2.79% ± 1.99% for PENELOPE and MCNP simulation, respectively. The differences between simulated FEPEs of two Monte Carlo codes were negligible except for low energies (< 100 keV).

Originele taal-2English
Artikelnummer109790
TijdschriftApplied Radiation and Isotopes
Volume174
DOI's
StatusPublished - aug. 2021

Vingerafdruk

Duik in de onderzoeksthema's van 'Optimization and validation of a LaBr3 (Ce) detector model for use in Monte Carlo simulations'. Samen vormen ze een unieke vingerafdruk.

Citeer dit