Samenvatting
Purpose or Objective
Surface guidance (SG) is used in combination with IGRT for patient positioning and monitoring during radiotherapy. In the case of intracranial treatments, especially stereotactic radiosurgery, the use of a closed immobilization mask is a common practice. These devices prevent patient motion but also hinder the task of tracking their external anatomy. The purpose of this study was to evaluate the effectiveness and the added value of surface tracking in detecting intra-fraction errors when immobilization devices are in place.
Material and Methods
Patient positioning information was obtained with stereoscopic x-rays (ExacTrac Dynamic, Brainlab, Munchen, Germany) taken when the gantry angle allowed image acquisition, providing 6DOF position information by comparing the x-ray with a set of DRR generated from the planning CT. The SG system obtains 6DOF information by continuous comparison of a ROI on the patient’s surface with a reference acquired after performing IGRT. In total, 173 snapshot verifications using stereoscopic X-rays and ST corresponding to 50 fractions of treatments including brain FSRT and SRS were analyzed. Discrepancies between errors measured by both systems were computed for all directions and all rotations.
Results
Mean absolute differences between x-rays and SG across all verifications were 0.13 mm (SD 0.24), -0.18 mm (SD 0.22) and -0.12 mm (SD 0.16) for the lateral, longitudinal and vertical directions, and -0.10° (SD 0.16), -0.07° (SD 0.13) and 0.09° (SD 0.15) for pitch, roll and yaw respectively. Absolute differences stayed below 0.4mm for all directions in 90% of the snapshots. Paired t-tests showed some correlation between the variables for lateral and vertical directions but no statistical significance for paired t-test or Wilcoxon tests was found for the data analyzed.
Surface guidance (SG) is used in combination with IGRT for patient positioning and monitoring during radiotherapy. In the case of intracranial treatments, especially stereotactic radiosurgery, the use of a closed immobilization mask is a common practice. These devices prevent patient motion but also hinder the task of tracking their external anatomy. The purpose of this study was to evaluate the effectiveness and the added value of surface tracking in detecting intra-fraction errors when immobilization devices are in place.
Material and Methods
Patient positioning information was obtained with stereoscopic x-rays (ExacTrac Dynamic, Brainlab, Munchen, Germany) taken when the gantry angle allowed image acquisition, providing 6DOF position information by comparing the x-ray with a set of DRR generated from the planning CT. The SG system obtains 6DOF information by continuous comparison of a ROI on the patient’s surface with a reference acquired after performing IGRT. In total, 173 snapshot verifications using stereoscopic X-rays and ST corresponding to 50 fractions of treatments including brain FSRT and SRS were analyzed. Discrepancies between errors measured by both systems were computed for all directions and all rotations.
Results
Mean absolute differences between x-rays and SG across all verifications were 0.13 mm (SD 0.24), -0.18 mm (SD 0.22) and -0.12 mm (SD 0.16) for the lateral, longitudinal and vertical directions, and -0.10° (SD 0.16), -0.07° (SD 0.13) and 0.09° (SD 0.15) for pitch, roll and yaw respectively. Absolute differences stayed below 0.4mm for all directions in 90% of the snapshots. Paired t-tests showed some correlation between the variables for lateral and vertical directions but no statistical significance for paired t-test or Wilcoxon tests was found for the data analyzed.
Originele taal-2 | English |
---|---|
Pagina's (van-tot) | S1614 - S1615 |
Aantal pagina's | 2 |
Tijdschrift | Radiotherapy and Oncology |
Volume | 182 |
DOI's | |
Status | Published - 2023 |
Evenement | ESTRO 2023 - Messe Wien Exhibition & Congress Center, Vienna, Austria Duur: 12 mei 2023 → 16 jun. 2023 https://www.estro.org/Congresses/ESTRO-2023 |