Introduction
To determine the monoamines dopamine (DA), noradrenaline (NAD) and serotonin (5-HT) in microdialysis samples, ion-pair liquid chromatography (LC) with electrochemical detection (ECD) is typically applied. Because of the low concentrations in the extracellular space (pM range), further dilution in the microdialysis samples, the complexity of the samples and the required temporal resolution, the analysis is challenging [1]. In addition, sensitivity should be further increased to allow detection of these monoamines in small brain areas and in smaller animals or to increase the validity in situations where monoamine levels are decreased. Furthermore, there is an increasing demand in the development of high-throughput analysis. In recent years, ultra-high performance LC (UHPLC) has been introduced as a way to increase the efficiency and speed of analysis by use of columns packed with sub-2 µm particles [2]. An optimal UHPLC-ECD method would be one with a short analysis time, a high sensitivity for the three monoamines and a good separation of NAD, which is the most hydrophilic compound, towards the front peak. The coupling of UHPLC with ECD for the analysis of low concentrated samples has not been explored till now.
Methods
The UHPLC-ECD method is based on the LC-ECD method described by Sarre et al [3]. The ALEXYS neurotransmitter analyzer consists of an AS 110 micro autosampler, an LC 110S pump (up to 1000 bar), an OR 110 degasser unit and the DECDADE II electrochemical detector, all provided by Antec (Zoeterwoude, The Netherlands). Instrument control and data acquisition are carried out by Clarity chromatography software of Data Apex (Prague, The Czech Republic). The mobile phase consists of 87% buffer solution at pH 5.5 and 13% acetonitrile. The composition of the buffer solution is 100 mM sodium acetate trihydrate, 20 mM citric acid monohydrate, 2 mM sodium decanesulfonate and 0.5 mM disodium edetate. Separations are performed on an Acquity UPLC BEH C8 column (100 x 1.0 mm, 1.7 µm) from Waters (Milford, Massachusetts, United States). The injection volume is 10 µL and the flow rate 60 µL/min, the column and detection temperature are set at 35 °C and the detection potential is + 450 mV vs. Ag/AgCl. Basal microdialysis samples are obtained from the hippocampus of the rat, using modified Ringer's solution as perfusion fluid.
Results and Discussion
The use of a column with smaller particles (1.7 µm instead of 5 µm) leads to higher peak efficiencies. This results in improved signal/noise ratios and detection limits below 100 pM for the three monoamines within 8 minutes analysis time.
Some issues need to be addressed to optimize the speed and sensitivity of the analysis. Firstly, lower detection limits are obtained by applying a low pass noise filter. However loss in peak shape, peak efficiency and resolution is observed. Fast Fourier transformation on the unfiltered chromatogram is more suitable, hereby achieving detection limits of approximately 50 pM for the three monoamines.
Secondly, increasing the mobile phase flow rate allows faster analysis. Moreover, following the Cottrell equation [4], high flow rates result in higher signals, what has been confirmed by scanning voltammograms. However, at flow rates above 100 µL/min, chromatographic splitting of the three monoamine peaks is observed. This is probably due to difficulties concerning the ion-pair formation between the monoamine and the ion-pair reagent at high flow rates. This phenomenon can be minimized by adding higher buffer concentrations (200 mM sodium acetate) to the mobile phase.
Thirdly, at a column temperature of 45 °C, a lower viscosity of the mobile phase occurs, providing lower backpressure and faster analysis. Since the column and the electrochemical cell are located in the same thermostat maintained compartment, the detection temperature is also 45 °C resulting in an overall higher sensitivity.
The separation of NAD towards other interfering peaks in the front remains however insufficient, especially in microdialysis samples. Adaptations to the mobile phase composition (percentage organic modifier, concentration ion-pair reagent) were not able to obtain an acceptable retention factor for NAD, in combination with a short retention time for 5-HT. Investigation of other column chemistries is part of ongoing research. With some improvements to the original method, it is although possible to determine dopamine and serotonin quantitatively within 6 minutes.
References
1. Sarre S, Michotte Y (2007) Liquid chromatographic methods used for microdialysis: an overview. In: Westerink BHC, Cremers TIFH. Handbook of Microdialysis. Vol 16. Elsevier, Amsterdam, pp 233-250.
2. Guillarme D, Ruta J, Rudaz S, Veuthey JL (2010) New trends in fast and high-resolution liquid chromatography: a critical comparison of existing approaches. Anal Bioanal Chem 397(3):1069-1082.
3. Sarre S, Thorré K, Smolders I, Michotte Y (1997) Microbore liquid chromatography analysis of monoamine transmitters. Methods Mol Biol 72:185-96.
4. Cottrell FG (1903) Residual current in galvanic polarization, regarded as a diffusion problem. Z. Phys. Chem 42:385-431.
To determine the monoamines dopamine (DA), noradrenaline (NAD) and serotonin (5-HT) in microdialysis samples, ion-pair liquid chromatography (LC) with electrochemical detection (ECD) is typically applied. Because of the low concentrations in the extracellular space (pM range), further dilution in the microdialysis samples, the complexity of the samples and the required temporal resolution, the analysis is challenging [1]. In addition, sensitivity should be further increased to allow detection of these monoamines in small brain areas and in smaller animals or to increase the validity in situations where monoamine levels are decreased. Furthermore, there is an increasing demand in the development of high-throughput analysis. In recent years, ultra-high performance LC (UHPLC) has been introduced as a way to increase the efficiency and speed of analysis by use of columns packed with sub-2 µm particles [2]. An optimal UHPLC-ECD method would be one with a short analysis time, a high sensitivity for the three monoamines and a good separation of NAD, which is the most hydrophilic compound, towards the front peak. The coupling of UHPLC with ECD for the analysis of low concentrated samples has not been explored till now.
Methods
The UHPLC-ECD method is based on the LC-ECD method described by Sarre et al [3]. The ALEXYS neurotransmitter analyzer consists of an AS 110 micro autosampler, an LC 110S pump (up to 1000 bar), an OR 110 degasser unit and the DECDADE II electrochemical detector, all provided by Antec (Zoeterwoude, The Netherlands). Instrument control and data acquisition are carried out by Clarity chromatography software of Data Apex (Prague, The Czech Republic). The mobile phase consists of 87% buffer solution at pH 5.5 and 13% acetonitrile. The composition of the buffer solution is 100 mM sodium acetate trihydrate, 20 mM citric acid monohydrate, 2 mM sodium decanesulfonate and 0.5 mM disodium edetate. Separations are performed on an Acquity UPLC BEH C8 column (100 x 1.0 mm, 1.7 µm) from Waters (Milford, Massachusetts, United States). The injection volume is 10 µL and the flow rate 60 µL/min, the column and detection temperature are set at 35 °C and the detection potential is + 450 mV vs. Ag/AgCl. Basal microdialysis samples are obtained from the hippocampus of the rat, using modified Ringer's solution as perfusion fluid.
Results and Discussion
The use of a column with smaller particles (1.7 µm instead of 5 µm) leads to higher peak efficiencies. This results in improved signal/noise ratios and detection limits below 100 pM for the three monoamines within 8 minutes analysis time.
Some issues need to be addressed to optimize the speed and sensitivity of the analysis. Firstly, lower detection limits are obtained by applying a low pass noise filter. However loss in peak shape, peak efficiency and resolution is observed. Fast Fourier transformation on the unfiltered chromatogram is more suitable, hereby achieving detection limits of approximately 50 pM for the three monoamines.
Secondly, increasing the mobile phase flow rate allows faster analysis. Moreover, following the Cottrell equation [4], high flow rates result in higher signals, what has been confirmed by scanning voltammograms. However, at flow rates above 100 µL/min, chromatographic splitting of the three monoamine peaks is observed. This is probably due to difficulties concerning the ion-pair formation between the monoamine and the ion-pair reagent at high flow rates. This phenomenon can be minimized by adding higher buffer concentrations (200 mM sodium acetate) to the mobile phase.
Thirdly, at a column temperature of 45 °C, a lower viscosity of the mobile phase occurs, providing lower backpressure and faster analysis. Since the column and the electrochemical cell are located in the same thermostat maintained compartment, the detection temperature is also 45 °C resulting in an overall higher sensitivity.
The separation of NAD towards other interfering peaks in the front remains however insufficient, especially in microdialysis samples. Adaptations to the mobile phase composition (percentage organic modifier, concentration ion-pair reagent) were not able to obtain an acceptable retention factor for NAD, in combination with a short retention time for 5-HT. Investigation of other column chemistries is part of ongoing research. With some improvements to the original method, it is although possible to determine dopamine and serotonin quantitatively within 6 minutes.
References
1. Sarre S, Michotte Y (2007) Liquid chromatographic methods used for microdialysis: an overview. In: Westerink BHC, Cremers TIFH. Handbook of Microdialysis. Vol 16. Elsevier, Amsterdam, pp 233-250.
2. Guillarme D, Ruta J, Rudaz S, Veuthey JL (2010) New trends in fast and high-resolution liquid chromatography: a critical comparison of existing approaches. Anal Bioanal Chem 397(3):1069-1082.
3. Sarre S, Thorré K, Smolders I, Michotte Y (1997) Microbore liquid chromatography analysis of monoamine transmitters. Methods Mol Biol 72:185-96.
4. Cottrell FG (1903) Residual current in galvanic polarization, regarded as a diffusion problem. Z. Phys. Chem 42:385-431.
| Originele taal-2 | English |
|---|---|
| Titel | Monitoring Molecules in Neuroscience, London |
| Status | Published - 21 sep. 2012 |
| Evenement | Unknown - Duur: 21 sep. 2012 → … |
Publicatie series
| Naam | Monitoring Molecules in Neuroscience, London |
|---|
Conference
| Conference | Unknown |
|---|---|
| Periode | 21/09/12 → … |
Vingerafdruk
Duik in de onderzoeksthema's van 'UHPLC with electrochemical detection for the quantitative determination of monoamines in microdialysis samples'. Samen vormen ze een unieke vingerafdruk.Activiteiten
- 1 Participation in conference
-
Monitoring Molecules in Neuroscience
Van Schoors, J. (Participant)
16 sep. 2012 → 21 sep. 2012Activiteit: Participation in conference
Citeer dit
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver