TY - JOUR
T1 - Validation of the Colibrí Instrument for Automated Preparation of MALDI-TOF MS Targets for Yeast Identification
AU - Heestermans, Robbe
AU - Herroelen, Pauline
AU - Emmerechts, Kristof
AU - Vandoorslaer, Kristof
AU - De Geyter, Deborah
AU - Demuyser, Thomas
AU - Wybo, Ingrid
AU - Pierard, Denis
AU - Muyldermans, Astrid
N1 - Publisher Copyright:
© 2022 Heestermans et al.
Copyright:
Copyright 2022 Elsevier B.V., All rights reserved.
PY - 2022/7/20
Y1 - 2022/7/20
N2 - Recently, Copan (Italy) introduced the Colibrí instrument for automated colony picking and preparation of matrix-assisted laser desorption ionization–time of flight (MALDI-TOF) target plates. Our study aimed to validate this system for yeasts as such testing has not been performed yet and is a missing link needed to implement the system for routine use. Fifty-five Candida strains were selected to evaluate the accuracy of Colibrí. For each strain, a sheep blood agar plate supplemented with X and V factors (HEM) and a Sabouraud agar plate (SAB) were inoculated and incubated using the WASPlab specimen processing system (Copan). After 18 h and 36 h of incubation, the isolates were spotted in parallel using Colibrí and manually onto MALDI-TOF target plates with the addition of formic acid and identified using MALDI-TOF mass spectrometry. The reproducibility was evaluated using ATCC reference and clinical isolate-derived strains. The cumulative percentage of acceptable identification scores (IDs) after 36 h was 91% for strains cultured on HEM plates using both Colibrí and the manual method. The SAB plates showed inferior results for both Colibrí (76%) and the manual method (78%). We observed an overall agreement of 92% at 18 h for identification of the strains on the HEM plates between Colibrí and the manual method and 94% after 36 h. For the SAB plates, the agreement was 78% after 18 h and 84% after 36 h. Apart from Candida dubliniensis and Candida tropicalis, all Candida species were identified with 100% accuracy using Colibrí on HEM plates. We observed good agreement between Colibrí and the manual reference method. These results demonstrate that Colibrí is a reliable system for MALDI-TOF target preparation for yeast identification, allowing increased standardization and less hands-on time.
AB - Recently, Copan (Italy) introduced the Colibrí instrument for automated colony picking and preparation of matrix-assisted laser desorption ionization–time of flight (MALDI-TOF) target plates. Our study aimed to validate this system for yeasts as such testing has not been performed yet and is a missing link needed to implement the system for routine use. Fifty-five Candida strains were selected to evaluate the accuracy of Colibrí. For each strain, a sheep blood agar plate supplemented with X and V factors (HEM) and a Sabouraud agar plate (SAB) were inoculated and incubated using the WASPlab specimen processing system (Copan). After 18 h and 36 h of incubation, the isolates were spotted in parallel using Colibrí and manually onto MALDI-TOF target plates with the addition of formic acid and identified using MALDI-TOF mass spectrometry. The reproducibility was evaluated using ATCC reference and clinical isolate-derived strains. The cumulative percentage of acceptable identification scores (IDs) after 36 h was 91% for strains cultured on HEM plates using both Colibrí and the manual method. The SAB plates showed inferior results for both Colibrí (76%) and the manual method (78%). We observed an overall agreement of 92% at 18 h for identification of the strains on the HEM plates between Colibrí and the manual method and 94% after 36 h. For the SAB plates, the agreement was 78% after 18 h and 84% after 36 h. Apart from Candida dubliniensis and Candida tropicalis, all Candida species were identified with 100% accuracy using Colibrí on HEM plates. We observed good agreement between Colibrí and the manual reference method. These results demonstrate that Colibrí is a reliable system for MALDI-TOF target preparation for yeast identification, allowing increased standardization and less hands-on time.
UR - http://www.scopus.com/inward/record.url?scp=85134854624&partnerID=8YFLogxK
U2 - 10.1128/jcm.00237-22
DO - 10.1128/jcm.00237-22
M3 - Article
VL - 60
SP - 1
EP - 7
JO - Journal of Clinical Microbiology
JF - Journal of Clinical Microbiology
SN - 0095-1137
IS - 7
ER -